Generally, POGO® cleaning is not recommended. However, in some cases the spring probe performance in relationship to electrical conductivity can be improved if the spring probe tips are cleaned of any contaminants. Contaminants can form an insulation barrier on the probe tip, thus reducing contact integrity.

One of the more widely used methods for cleaning spring contact probes involves the use of brushes to clean the probe heads without probe removal from the test fixture. This technique allows for more frequent maintenance resulting in improved fixture reliability. After brushing contaminants free from the probes, the fixture should be vacuumed to insure no remaining particles create future problems.

Another cleaning method involves removal of probes from the test fixture, bundling them together, and submerging only the probe tips in a shallow pan of safe solution such as alcohol or citric cleaner for five minutes. After soaking, the probe tips can be scrubbed with a soft bristle brush to remove any residue still remaining, then rinsed and dried. The probes can then be installed back into the test fixture. We caution the reader that this cleaning method should be attempted only as a last resort, as cleaning fluids and solvents can wash contaminants into the probes as well as the fixture.

For more information on probe maintenance, contact our professional sales staff.